SUPPLEMENTAL
DOCUMENT SD-2
FOR PART
IVB
QUALITY
ASSURANCE/VALIDATION OF ANALYTICAL METHODS
Contents Page
Preparing Validation Plans
Preface
. 1
Section I Analytical
Techniques Elements to Consider
Introduction
.....
2
1 Infrared Spectroscopy (IR)
... 3
2 Mass Spectrometry (MS)
.. 6
3 Nuclear Magnetic
Resonance Spectroscopy (NMR)...
9
4 Raman Spectroscopy
...
.... 12
5 Capillary Electrophoresis (CE)
. 15
6 Gas Chromatography (GC)
.. 18
7 Ion Mobility
Spectrometry (IMS)
..
21
8 High
Performance Liquid Chromatography (HPLC)
.
24
10 Thin Layer Chromatography
(TLC)
.
30
11 Color
Tests
.
. 33
12 Fluorescence
Spectrophotometry
.
.. 35
13 Immunoassay
. 38
14 Melting Point
..
. 41
15 Ultraviolet Spectrophotometry
(UV)
44
Section II A Completed Validation Plan
Introduction..
...
47
Example
.
.. 47
SUPPLEMENTAL
DOCUMENT SD-2
FOR PART
IVB
QUALITY
ASSURANCE/VALIDATION OF ANALYTICAL METHODS
PREPARING VALIDATION PLANS
This supplemental document is designed to assist
laboratories develop a general validation plan which meets their individual
requirements. The document is intended
to be used in conjunction with SWGDRUG Recommendations,
Section 2, Part IVB.
The supplemental document consists of two
sections. Section I provides guidance on the issues to consider when
using various analytical techniques. Section II is an example of a completed validation plan.
Note In
completing the validation process all sections of SWGDRUG Recommendations, Part
IVB need to be considered.
Section I: Analytical Techniques Elements to
Consider
This
section details technique specific properties including technique strengths
and technique limitations that may affect the design of a validation
plan. The analytical techniques
described correspond to those in categories A, B and C of the Part
Table 1: Categories of Analytical Techniques
|
Category A |
Category B |
Category C |
|
Infrared Spectroscopy |
Capillary Electrophoresis |
Color Tests |
|
Mass Spectrometry |
Gas Chromatography |
Fluorescence Spectroscopy |
|
Nuclear Magnetic Resonance
Spectroscopy |
Ion Mobility Spectrometry |
Immunoassay |
|
Raman Spectroscopy |
Liquid Chromatography |
Melting Point |
|
|
Microcrystalline tests |
Ultraviolet Spectroscopy |
|
|
Pharmaceutical Identifiers |
|
|
|
Thin Layer Chromatography |
|
|
|
Cannabis only: Macroscopic Examination Microscopic Examination |
|
Reference:
·
EAL-P11 European
Cooperation for Accreditation of Laboratories
·
ILAC Guidelines
for Forensic Laboratories Feb 2001, 5.4.5.1
·
Eurachem, The Fitness for Purpose of Analytical
Methods, 1998
·
Federal Register
Vol. 60 no. 40 pg 11259,
1.1 Technique
Strengths
Samples can be recovered for additional tests.
IR
generates the highest discriminating capability. It may discriminate between diastereomers
(pseudoephedrine/ephedrine) and free base/acid and salt forms.
1.2 Technique
Limitations
Pure samples may give different spectra due to
polymorphism.
Chemical
composition should not change during the analysis. For example, care must be taken to address
volatility, heat, and pressure effects.
1.3 Purpose/Scope
IR
yields structural information that will provide sufficient selectivity that
generates the highest discriminating capability (category
A).
1.4 Analytical
Method
1.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
1.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List the instrument conditions
1.5 Reference
Materials
Utilize a polystyrene film and compare to a polystyrene
standard spectrum.
Repeat
this process utilizing a commonly encountered drug standard suitable for this
method.
1.6 Performance
Characteristics
1.6.1 Selectivity
For
determination of closely related compounds, standards of each should be tested
on the system to show selectivity.
IR
may discriminate between diastereoisomers (pseudoephedrine/ephedrine) and free
base/acid and salt forms.
However, IR cannot distinguish enantiomers.
1.6.2 Matrix
Effects
Samples need to be dry to minimize water interferences.
Address
the possibility of ion exchange (alkali halides such as KCl and KBr) during
sample preparation.
Analytes commonly require purification sufficient for
their intended purpose.
1.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
1.6.4 Accuracy
1.6.4.1 Precision
(Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
1.6.4.2 Trueness
Trueness must be determined for
quantitative methods.
1.6.5 Range
Limit
of Detection (
◦ Determine
this through measuring the response of different amounts of analyte.
◦ For
most instruments this is in the microgram range.
Limit of quantitation (LOQ) should be determined.
Linearity must be determined for all quantitative methods.
1.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for the
identification (e.g., wavenumber resolution, concentration, humidity, temperature).
1.6.7 Ruggedness
Ruggedness
may be determined for qualitative or quantitative methods. Alter the analysts, instrumentation and
environment and assess the changes in accuracy.
1.7 Uncertainty
The uncertainty of the method must be assessed for
quantitative methods.
1.8 Quality
control
1.9 Reference
1.9.1 EAL-P11 European Cooperation for
Accreditation of Laboratories
1.9.2 ILAC Guidelines for Forensic Laboratories,
Feb 2001, 5.4.5.1
1.9.3 Eurachem, The
Fitness for Purpose of Analytical Methods, 1998
2.1 Technique
Strengths
The technique may discriminate between diastereomers.
Mass
spectra can be interpreted to aid in characterizing an unknown drug through
structural elucidation.
Techniques to interface MS with GC, LC and CE are readily
available.
Different
ionization techniques enable MS analysis of stabile/labile and polar/non-polar
compounds.
2.2 Technique
Limitations
Mass Spectrometry cannot discriminate enantiomers.
Mass
Spectrometry cannot be used to identify salt forms nor determine if a salt or
free drug is present.
Stability
of measured compounds: Fragmentation of
some drugs may occur leaving no molecular ion (certain barbiturates), or
similar patterns (e.g., Bufotenine, Psilocyn, Psilocybin).
2.3 Purpose/Scope
A
mass spectrum yields structural information which may provide sufficient
selectivity to allow for the highest discriminating capability (category A).
When
used in combination with gas or liquid chromatography, several compounds
present in the same sample can be identified and quantified. The same applies to the multidimensional MS
techniques.
2.4 Analytical
Method
2.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
2.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List the instrument conditions.
2.5 Reference
Materials
Utilize a standard calibration compound such as PFTB
◦ Acquire
a mass spectrum of this compound and compare it to a standard spectrum.
◦ Repeat
this process utilizing a commonly encountered drug standard suitable for this
method.
2.6 Performance
Characteristics
2.6.1 Selectivity
For
determination of closely related compounds, standards of each should be tested
on the system to show selectivity.
2.6.2 Matrix
effects
Co-elution and a high concentration of substance can cause
a matrix effect.
2.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
2.6.4 Accuracy
2.6.4.1 Precision
(Repeatability/Reproducibility):
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
2.6.4.2 Trueness:
Must be determined for quantitative methods.
2.6.5 Range
2.6.5.1 Limit
of Detection (
Select
the criteria for the mass spectrum below which no identification will be made.
Determine
the limit of detection by measuring the response of different amounts of
analyte.
For most instruments the limit of detection is in the
picogram to nanogram range.
2.6.5.2 Limit
of quantitation (LOQ):
LOQ must be determined for all quantitative methods.
Determine the lowest concentration that has an acceptable
level of uncertainty.
Concentration
ranges should be in the order of published spectra to avoid difficulties with
comparison. For example, high analyte
concentration in the sample preparation may cause variations in m/e ratios. The response will differ between instruments
and analytes.
Linearity must be determined for all quantitative methods.
2.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for the
identification (e.g., concentration, humidity, temperature).
2.6.7 Ruggedness
Ruggedness may be determined for qualitative and
quantitative methods.
2.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
2.8 Quality Control
2.9 Reference
3 NUCLEAR
MAGNETIC RESONANCE SPECTROSCOPY (NMR)
3.1
Technique Strengths
·
Methods
developed on one NMR are conveyable to any other NMR so long as the following
items are the same: magnetic field
strength (the same or greater), and probe linearity and uniform response are
checked on one drug. This is due to the
inherent stability of the instrument and the manner in which it detects
compounds.
·
Samples may be
recovered for further analysis.
·
Multiple
experiments can be run on one sample depending on the capabilities of the
specific instrument.
·
Specific
structural information may be obtained from spectra of several nuclei (i.e.,
hydrogen-1 or protium, carbon-13, nitrogen-15, etc.).
·
Enhanced
selectivity may be achieved by using various one or multi-dimensional analysis
techniques, chemical exchange, or adding a shift reagent.
·
Enantiomers
can be differentiated.
·
Thermally
unstable drugs can be analyzed without decomposition.
·
Target compounds
can be analyzed without derivatization.
·
Multiple
solvents are available to enhance selectivity and/or solubility.
·
Unlike
UV, FID, or MS detectors, NMR response is based on the molar quantity of nuclei
at a given frequency and is the same for all compounds.
·
Quantitation
is performed without the use of a reference drug standard of the target
compound.
·
The
technique can enable simultaneous identification and quantitation.
·
Sample
run times, sample to sample, are short.
3.2
Technique
Limitations
·
·
Solvents can
interfere with the peaks of the sample being analyzed. Solvents usually used for proton NMR have
deuterium (2H) substituted for protons (1H).
·
Very complex
mixtures can lead to the absence of clean well resolved signals to integrate
making integration and quantitation difficult.
3.3 Purpose/Scope
The
NMR instrument can be used for both qualitative and quantitative analyses. It allows identification and structure
elucidation of an analyte that will provide sufficient selectivity to generate
the highest discriminating capability (category A).
3.4 Analytical Method
3.4.1 Sample preparation
List
the required sample preparation schemes, including the solvent appropriate for
the nucleus being monitored.
3.4.2 Instrumental
parameters
·
Identify
the instrument and equipment utilized.
·
List
instrument conditions.
3.5
Reference Materials
·
The referencing
of the chemical shift requires either the use of tetramethylsilane (
·
Ethylbenzene can
be utilized to check the calibration of the chemical shift and to demonstrate
appropriate resolution.
·
For
quantitation, internal standards must be of high purity, non-reactive, soluble
in the solvent and have chemical shifts that do not interfere with compounds
that will be encountered in the sample (e.g., benzoic acid).
3.6 Performance Characteristics
3.6.1
Selectivity
·
For determination
of closely related compounds, standards of each should be tested on the system
to show selectivity.
·
In cases of
signal overlap, the interfering compounds contribution can be determined and
subtracted from the mixed integral to obtain the target compounds integral
value.
3.6.2
Matrix Effects
The internal standard and
the analyte should be stable and fully soluble in the selected NMR
solvent. The NMR sample should be free
of particulate matter.
3.6.3
Recovery
Sample
recovery may be determined for quantitative analysis.
3.6.4
Accuracy
·
Precision
(Repeatability/Reproducibility): Demonstrate the reproducibility of the
instrument by running a reference material a minimum of 10 times.
·
Trueness: Must
be determined for quantitative methods.
3.6.5
Range
·
Limit of Detection
(
·
Limit of
quantitation (LOQ): must be determined for all quantitative methods. Determine
the lowest concentration that has an acceptable level of uncertainty. Linearity
of the probe based on concentration must be determined for quantitation. A probe that is linear in one method will be
linear in all methods.
3.6.6
Robustness
It is not applicable for
this technique. The method establishes
the instruments parameters; they are not allowed to be changed.
3.6.7
Ruggedness
Not applicable for this
instrument.
3.7
Uncertainty
Uncertainty should be
evaluated for quantitative methods.
3.8 Quality control
3.9 Reference
4.1 Technique Strengths
Raman
generates a very high discriminating capability unaffected by glass or plastic
containers.
Little to no sample preparation is required.
It is compatible with remote sampling and fiber optics.
4.2 Technique Limitations
Raman
needs a fairly concentrated sample and may not be suitable for residue
analysis.
Instrumental
effects can be subtle and difficult to understand and control (for example,
wavelength-dependent changes in the solid angle of the collected Raman light
arising from changing indices of refraction).
4.3 Purpose/Scope
Raman
spectroscopy yields structural information that will provide sufficient
selectivity that generates the highest discriminating capability (category A).
4.4 Analytical Method
Identify the procedures to be utilized in the validation
process.
Verification of correct x- and y-axis calibration is
required.
4.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
4.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List the instrument conditions.
4.5 Reference Materials
A compound must have a Raman-active vibrational mode.
A
series of neat organic liquids with published peak positions of the Raman
spectra can be used for x-axis calibration validation.
Repeat
this process utilizing a commonly encountered drug suitable for this method.
4.6 Performance
Characteristics
4.6.1 Selectivity
Raman
generates a very high discriminating capability unaffected by glass or plastic
containers.
This
technique is highly selective, for example isomers may be detected from the
changes in the molecular vibrational frequencies.
For
determination of closely related compounds, standards of each should be tested
on the system to show selectivity.
4.6.2 Matrix
Effects
Fluorescence can swamp the Raman signal.
Compounds in aqueous solution are easily measured.
4.6.3 Recovery
Sample recovery may be determined for quantitative analysis.
4.6.4 Accuracy
4.6.4.1 Precision
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times
4.6.4.2 Trueness
Trueness must be determined for quantitative methods.
4.6.5 Range
4.6.5.1 Limit
of Detection
A
peak-to-peak ratio should be determined below which no identification will be
made.
Determine
the limit of detection by measuring the response of different amounts of
analyte.
4.6.5.2 Limit
of Quantitation (LOQ)
4.6.5.3 Linearity
4.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for the
identification (e.g., scan time, resolution).
4.6.7 Ruggedness
4.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
4.8 Quality
Control
4.9 Reference
4.9.1 American
Society for Testing and Materials, Standard Guide for Raman Shift Standards for
Spectrometer Calibration, Standard E-1840-96, 1998.
4.9.2 R.L.
McCreery, Raman Spectroscopy for Chemical Analysis, vol. 157 of Chemical
Analysis, J.D. Winefordner, ed.,
5 CAPILLARY
ELECTROPHORESIS (CE)
5.1 Technique
Strengths
CE
provides high speed, high-resolution separations on small sample volumes (0.1nL
to 10mL).
A
variety of detection methods can be used, to include fluorescence, absorbance,
electrochemical, and mass spectrometry detectors.
Potentials
up to 60,000V can be safely applied, allowing increases in CEs speed and
resolution.
CE
employs electro-osmotic flow.
Electro-osmotic flow creates solution flow with a flat profile, as
opposed to the parabolic profile created by liquid chromatography. The flat solution profile doesnt contribute
significantly to band broadening.
CE
allows the user to reverse the direction of normal electro-osmotic flow, which
speeds up the separation of anions.
CE
works quite well with compounds that will not separate by gas chromatography
because they are; too polar, thermally labile, or nonvolatile.
A
chiral buffer allows for the separation of optical isomers.
5.2 Technique
Limitations
Long migration times may have greater variability within
the peak area.
Reproducibility of migration times is less reproducible
than in GC
5.3 Purpose/Scope
Capillary
electrophoresis is a high-speed, high-resolution separation process that can be
used for qualitative and quantitative analysis and for separation of chiral
pairs of drugs.
5.4 Analytical
Method
5.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
5.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List
the instrumental conditions such as capillary temperature and specifications,
voltage ramp, injection times, and buffer.
5.5 Reference
Materials
A reference
material or mixtures of reference materials of
the drugs to be analyzed are suitable for method validation.
Mixtures
may include a standard of the drug, common additives, and drugs similar to the
analyte.
5.6 Performance Characteristics
5.6.1 Selectivity
During
separation, CE provides various means to adjust the α values thus giving
good resolution for the target compounds in most applications.
Evaluate
the selectivity by using a representative number of drugs and potential
adulterants/diluents.
5.6.2 Matrix
Effects
Samples must be carefully filtered.
5.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
5.6.4 Accuracy
5.6.4.1 Precision
(Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum of 10 times.
5.6.4.2 Trueness
Trueness must be determined for quantitative methods.
5.6.5 Range
5.6.5.1 Limit
of Detection (
CE
can be a very sensitive technique.
Determine the
5.6.5.2 Limit
of Quantitation (LOQ)
5.6.5.3 Linearity
5.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for a
suitable comparison.
5.6.7 Ruggedness
5.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
5.8 Quality
Control
5.9 Reference
6.1 Technique
Strengths
Capillary columns provide many theoretical plates.
Detector response is proportional to sample concentration.
GC demonstrates a high degree of selectivity
Enantiomers
can be determined using properly validated chiral columns or derivatization
techniques.
6.2 Technique
Limitations
Although
highly selective, the possibility exists that another compound will elute at
the same retention time.
Salts
are usually dissociated during the injection process and cannot be identified.
Some
salt forms will cause excessive tailing and should be extracted prior to
injection.
Chemical decomposition can occur in the injector port or
during the analysis.
Samples must be capable of volatilization.
6.3 Purpose/Scope
Gas
Chromatography is a separation and comparison technique that will provide data
that can indicate the probable identity of the analyte and the possible
presence of additional sample components.
It
can be used as a quantitative method.
6.4 Analytical
Method
6.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
6.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List
the instrumental conditions, to include, injector and detector temperature,
column temperature and ramp (if appropriate), mobile phase.
6.5 Reference
Materials
A reference material or mixtures of reference
materials of the drugs to be analyzed are suitable for method validations.
Mixtures
may include a standard of the drug, common additives, and drugs similar to the
analyte.
6.6 Performance
Characteristics
6.6.1 Selectivity
Gas Chromatography possesses moderate discriminatory
power.
For
determination of closely related compounds, standards of each should be tested
on the system to show selectivity.
6.6.2 Matrix
Effects
Determine
the common excipients, additives or solvents that may react with the analyte in
the GC.
6.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
6.6.4 Accuracy
6.6.4.1 Precision
(Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
6.6.4.2 Trueness
Trueness must be determined for quantitative methods.
6.6.5 Range
6.6.5.1 Limit
of Detection
Gas
chromatography is a sensitive technique and is dependent upon the
chromatographic system used and the analyte present.
Determine
the sensitivity by measuring the response of different amounts of analyte.
Typical
sensitivity is on the order of picogram to nanogram range.
6.6.5.2 Limit
of Quantitation (LOQ)
6.6.5.3 Linearity
6.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for a
suitable comparison.
6.6.7 Ruggedness
6.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
6.8 Quality
Control
6.9 Reference
7
7.1 Technique
Strengths
IMS
instruments are relatively small and may be utilized in the field to
presumptively screen drugs.
Generally, results can be obtained in less than one
minute.
Analytes are detectable in the nanogram range.
A
properly obtained IMS plasmagram provides the presumptive identification of
drugs.
7.2 Technique
Limitations
IMS is not a specific identification technique.
Drugs may have similar drift times
IMS is a destructive technique
Concentration
of reference material and unknowns should not be so high as to saturate the
instrument. For example, high analyte
concentration may change the drift time.
The
concentration ranges should be determined by experiment to identify the
effective range.
Instrument
is sensitive to temperature fluctuation and changes in atmospheric pressure.
7.3 Purpose/Scope
Ion
Mobility Spectrometry refers to the principles, practice, and instrumentation
for characterizing chemical substances by measurement of their gas-phase ion
mobilities.
This
analytical technique may provide presumptive identification of drugs.
7.4 Analytical
Method
7.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
7.4.2 Instrumental
parameters
Identify
the instrument and equipment utilized, to include the sample collection
equipment and sample collection method.
List the instrumental conditions.
7.5 Reference
Materials
Utilize the internal calibrant recommended by the
instrument manufacturer.
Utilize
standard external calibrants as references such as cocaine or methamphetamine
for the target compounds.
7.6 Performance
Characteristics
7.6.1 Selectivity
For
determination of closely related compounds, standards of each as well as a mixture
should be tested on the system to show selectivity.
7.6.2 Matrix
Effects
Dirt,
hair, or fibers collected in the sampling device may prevent the desorption of
the analyte.
7.6.3 Recovery
Sample recovery may be determined for quantitative analysis.
7.6.4 Accuracy
7.6.4.1 Precision
(Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
7.6.4.2 Trueness
Trueness must be determined for quantitative methods.
7.6.5 Range
7.6.5.1 Limit
of Detection
Instruments may have different detection limits.
Determine
the limit of detection by measuring the response of different amounts of target
analytes.
7.6.5.2 Limit
of Quantitation
7.6.5.3 Linearity
7.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for a
suitable comparison.
7.6.7 Ruggedness
7.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
7.8 Quality
Control
7.9 Reference
7.9.1 Eiceman, Karpas, Ion Mobility Spectrometry,
8 HIGH
PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)
8.1 Technique
Strengths
Non-destructive, samples can be recovered for additional
tests.
Thermally labile drugs can be analyzed without
decomposition.
Non-volatile drugs can be analyzed without derivatization.
HPLC can be a screening tool for certain groups or
compounds.
8.2 Technique
Limitations
The highest purity solvents available should be used.
As peak symmetry decreases, integration becomes less
reliable.
System
should be allowed to equilibrate before samples are run in order to assure
reproducible conditions.
An appropriate standard must be included with each set of
samples.
Potential carryover must be taken into consideration.
8.3 Purpose/Scope
HPLC
is a separation and comparison technique that provides data that can indicate
the probable identity of the analyte and the possible presence of additional
sample components.
It
can be used as a quantitative method, combined with various detectors for
greater selectivity, used for preparative purposes or used to separate
enantiomers by utilizing chiral columns.
8.4 Analytical
Method
8.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
8.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List
the instrumental conditions, such as elution time, temperature, flow rates and
detector settings.
8.5 Reference
Materials
Reference
materials or a mixture of reference materials of the drugs to be analyzed are
suitable for method validation.
Mixtures
may include a standard of the drug, internal standards, common additives, and
drugs similar to the analyte.
8.6 Performance
Characteristics
8.6.1 Selectivity
HPLC possesses moderate discriminatory power.
Selectivity can be enhanced with the use of different
detectors.
For
determination of closely related compounds, standards of each should be tested
on the system to show selectivity.
8.6.2 Matrix
Effects
Compounds
other than the analyte may impede the progress of the analyte through the
system.
The
solvent containing the analyte may require a solvent system of a similar
strength.
8.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
8.6.4 Accuracy
8.6.4.1 Precision
(Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
8.6.4.2 Trueness
Trueness must be determined for quantitative methods.
8.6.5 Range
8.6.5.1 Limit
of detection
HPLC can be a sensitive technique.
Determine
the sensitivity by measuring the response of different amounts of analyte.
For most systems this is in the microgram or submicrogram
range.
8.6.5.2 Limit
of quantitation
8.6.5.3 Linearity
8.6.6 Robustness
8.6.7 Ruggedness
8.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
8.8 Quality
Control
8.9 Reference
8.9.1 The
9.1 Technique
Strengths
Most
crystals formed are temporary complexes and the test compounds are recoverable
from the test slide.
Individual
tests require only simple glass slides and one or two drops of the crystal
reagent.
Crystal
tests adopted by laboratories form quickly and are easily read or they are not
incorporated into the analytical scheme.
Many
habits of crystals differ significantly from each other and are easily
described.
Closely related analogs may be readily differentiated.
9.2 Technique
Limitations
Slower
forming crystals may be due to the reagent drying and will form on the edges of
the solution.
Habits often change with continued crystal growth.
Relatively
large sample amounts are required to obtain crystals [usually several
milligrams].
Reviewable
data must be produced through observation by an additional analyst or the
crystals must be photographed/imaged before overgrowth occurs.
9.3 Purpose/Scope
Microcrystalline
tests can be used for determining the presence of many chemicals including both
controlled substances and other related compounds.
Microcrystals
form readily from the combination of many controlled substances and specific
reagents and are recognized by their visual characteristics (habits) to the
trained analyst.
9.4 Analytical
Method
9.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment.
9.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
Identify
the procedures to be utilized. Provide
the necessary documentation such as the techniques used to induce microcrystal
growth with the substance.
List the instrumental conditions.
9.5 Reference
Materials
Standard
reference material samples of the compounds to be validated as well as closely
related structures should be examined.
9.6 Performance
Characteristics
9.6.1 Selectivity
Known
analogs of the desired compound, as well as common diluents should be examined
to verify that the selectivity of the reagent is adequate.
9.6.2 Matrix
Effects
Either the solvent or other compounds may limit crystal
formation.
High ambient temperatures may reduce crystallization.
9.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
9.6.4 Accuracy
9.6.4.1 Precision
(Repeatability/Reproducibility)
A
series of samples should be examined under differing temperature and humidity
conditions.
9.6.4.2 Trueness
Trueness must be determined for quantitative methods.
9.6.5 Range
9.6.5.1 Limit
of Detection
Known
solution strengths should be tested with each of the common diluents to
establish the limit of detection.
9.6.5.2 Limit
of Quantitation
9.6.5.3 Linearity
9.6.6 Robustness
9.6.7 Ruggedness
9.7 Uncertainty
9.8 Quality
Control
9.9 Reference
9.9.1 Clarke,
E.G.C., Isolation and Identification of Drugs, Volume 1,
9.9.2 Fulton,
Charles C., Modern Microcrystal Tests for Drugs,
10
10.1 Technique
Strengths
Samples
can be recovered for additional tests if non-destructive visualization
techniques are employed.
Multiple samples can be spotted on the same plate.
Exposure
of an eluted and dried plate to iodine vapor will, in general, visualize the
drug of interest.
Selection
of a specific eluent can increase the selectivity of the system for isolation
of the targeted compound.
Selectivity can also be increased by multiple developments
on a single plate.
An appropriate standard must be included with each
analysis.
If a
mixture of standards displaying adequate separation is included,
self-verification is provided.
10.2 Technique
Limitations
The
amount of analyte spotted on a TLC plate should be sufficient for the intended
use.
If
comparison to a standard is being made, the amounts of sample and standard
spotted should be similar.
Edge
effects result from the eluent evaporation off of the edges of the plate and
inequalities in thickness and density of the stationary phase at the edge of
the plate.
Edge
effects may result in analyte migration toward the edge of the plate and
non-circular spot shape.
Salt
forms can also affect spot shape and Rf values.
For example, cocaine hydrochloride usually tails more than cocaine base.
Chemical
composition of the analyte should not change during TLC. The analyte should be stable in the eluent
10.3 Purpose/Scope
Thin-layer
chromatography is a quick separation and comparison technique that will provide
data that can indicate the probable identity of the analyte and the possible
presence of additional sample components.
It
can be used as a semi-quantitative method, be combined with degradation methods
for greater selectivity, or be used as a preparative method.
10.4 Analytical
Method
10.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected equipment.
10.4.2 Instrumental
parameters
Identify the equipment utilized.
Identify
the procedures to be utilized. Provide
the necessary documentation regarding solvent systems, and visualization
techniques.
Provide
the necessary documentation such as Rf values and description, such as the
color and shape of visualized analytes.
10.5 Reference
Materials
A reference
material or mixtures of reference materials of the drugs to be analyzed are
suitable for method validation.
Mixtures
may include a standard of the drug, common additives, and drugs similar to the
analyte.
10.6 Performance
Characteristics
10.6.1 Selectivity
TLC possesses moderate selectivity.
A
match of Rf between two spots only means that the two compounds have some
probability of being identical in composition.
Selectivity
may be enhanced by the use of different visualization techniques.
10.6.2 Matrix
Effects
Oils
and very concentrated co-eluting compounds can affect the Rf of the drug of
interest.
10.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
10.6.4 Accuracy
10.6.4.1 Precision (Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
10.6.4.2 Trueness
Trueness must be determined for quantitative methods.
10.6.5 Range
10.6.5.1 Limit of detection
The
limit of detection of TLC is very dependent on the nature of the analyte and
the selected detection method.
Determine
the sensitivity by measuring the response of different amounts of analyte.
10.6.5.2 Limit of quantitation
10.6.5.3 Linearity
10.6.6 Robustness
10.6.7 Ruggedness
10.7
Uncertainty
10.8
Quality Control
10.9
Reference
11.1 Technique
Strengths
Inexpensive: The equipment needed, a test tube or
multi-well porcelain spot plate and a dropping bottle, is not expensive.
Speed: If a color is to be developed by the sample
and reagent, it will happen within a characteristic short time (usually less
than a minute).
Multitasking: A
large number of samples may be tested simultaneously.
11.2 Technique
Limitations
Drugs
with similar structure may give the same colors. For example, dextropropoxyphene can give the
same color changes as cocaine in the Scott test.
Some
color test reagents consist of chemicals that are inherently dangerous. For example, the Marquis reagent contains
concentrated sulfuric acid. This makes
wearing of eye protection very important while using the Marquis reagent.
Colors
developed after a lengthy exposure of the sample to the reagent are not
reliable. For example, the sulfuric acid
in the Marquis reagent will decompose almost any drug over time. Therefore, the brown color developed by the
sample in Marquis solution over ten or twenty minutes cannot be taken as an
indication of the presence of methamphetamine in the sample.
11.3 Purpose/Scope
Color
tests are used as preliminary tests to indicate that a certain drug may or may
not be present in an unextracted sample.
A
positive result does not indicate that a specific drug is present, but it does
indicate that a certain class of drug is present.
The
result of the color test depends on the reaction of a certain moiety of the
drug molecule with the color test reagent which is characteristic of the sample
and causes a color change.
Since
the results are detected visually, care must be taken that the analyst be
thoroughly tested for the visual ability to detect very slight color changes.
11.4 Analytical
Method
11.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected equipment.
11.4.2 Instrumental
parameters
11.5 Reference
Materials
Reaction
of a standard compound with a color reagent to give the expected color will
serve as a validation test of that color reagent.
11.6 Performance
Characteristics
11.6.1 Selectivity
For
analysis of closely related compounds, standards of each should be tested using
the color test reagent to show selectivity.
11.6.2 Matrix
Effects
11.6.3 Recovery
11.6.4 Accuracy
11.6.4.1 Precision (Repeatability/Reproducibility)
Demonstrate
the reproducibility by running a reference material a minimum of 10 times.
11.6.4.2 Trueness
11.6.5 Range
11.6.5.1 Limit of Detection
Determine
the limit of detection by measuring the response of different amounts of
analyte.
11.6.5.2 Limit of quantitation
11.6.5.3 Linearity
11.6.6 Robustness
11.6.7 Ruggedness
11.7 Uncertainty
11.8 Quality
Control
11.9 Reference
12 FLUORESCENCE
SPECTROPHOTOMETRY
12.1 Technique
Strengths
Samples can be recovered for additional tests.
Identification: The fluorescence spectrum of an unknown pure
analyte when compared to a known standard can provide preliminary
identification of the compound.
The
fluorescence spectrum of an unknown pure analyte can provide information
concerning chromophores present in the analyte.
Can be used as a screening method for unknown drug
compounds.
Pure
analytes or analytes showing no interference are suitable for quantitative
analysis.
Appropriate
standards should be run to demonstrate reproducibility of the procedure.
12.2 Technique Limitations
Concentration ranges should be sufficient for the intended
use.
Although
selective, many compounds contain the same chromophores that contribute to the
data received. Under normal conditions,
these appear identical.
The
presence of the analyte as a salt will effect its solubility in a given
solvent, however, the type of salt cannot be determined, and once dissolved,
the analyte and its salt will give the same response.
Under normal conditions the compounds are stable.
Not all compounds show characteristic fluorescence spectra
Phosphorescence,
solvent fluorescence, matrix fluorescence, and light scattering can affect
results.
12.3 Purpose/Scope
A
fluorescence spectrum of an unknown analyte can give a preliminary
identification as to what compound may be present by comparison of the data
received to that of a standard run under the same conditions.
It
can be used as a quantitative method.
12.4 Analytical
Method
12.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment
12.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List
instrument conditions to include: excitation wavelength, absorbance wavelength
and solvents.
12.5 Reference
Materials
Reference
materials of the drugs to be analyzed are suitable for method validation.
Mixtures
may include a standard of the drug, common additives, and drugs similar to the
analyte may also be used
12.6 Performance
Characteristics
12.6.1 Selectivity
Fluorescence
spectra possess limited discriminatory power.
A match of the emission spectra between a sample analyte and a known
standard may mean that the two are identical.
However, in reality, it means that the two may possess the same types of
chromophore and respond the same under the conditions used.
Standards
must be run frequently to insure method and instrument stability.
It
provides a useful screening method, or if used in a proper scheme, a
confirmation of previously identified substances.
12.6.2 Matrix
effects
12.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
12.6.4 Accuracy
12.6.4.1 Precision (Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
12.6.4.2 Trueness
Trueness must be determined for quantitative methods.
12.6.5 Range
12.6.5.1 Limit of detection
Fluorescence
spectrophotometry is a sensitive technique with selective compounds. It is dependent upon the compound of
interest.
12.6.5.2 Limit of quantification
12.6.5.3 Linearity
12.6.6
Robustness
Determine
the amount of change to instrumental parameters that will still allow for a
suitable comparison.
12.6.7 Ruggedness
12.7 Uncertainty
12.8 Quality
Control
12.9 Reference
13.1 Technique
Strengths
Many labels available for detection: radionucleotides, enzymes, fluorescence
Need
to establish the limits of detection for the controlled substance and
cross-reacting compounds.
Suitable for manual or automated batch analyses
Procedure
should be designed with included controls to demonstrate that the method is
free of carryover.
13.2 Technique
Limitations
Majority
of reagents have cross reactivities.
Some stereo specific reagents exist with stereoisomers exhibiting less
activity than compound that is being sought.
The shape of the antigenic site on the antibody controls this.
Concentration
ranges should be sufficient for the intended use. Commercial kits are designed for
concentrations appropriate for toxicology samples. To be used as semi-quantitative analyses,
appropriate range standards must be included.
13.3 Purpose/Scope
Immunoassays
can determine the probable identity of several different drug classes and are
suitable for semi-quantitative analysis for compounds included within the
tested class.
13.4 Analytical
Method
13.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment
13.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
List instrumental conditions
13.5 Reference
Materials
Reference
materials of the drugs to be analyzed are suitable for method validation.
Study
should also include any compounds reported by the manufacturer as cross-reacting
species.
Compounds,
which are structurally similar, should also be examined even if no previous
cross-reactions have been reported.
13.6 Performance
Characteristics
13.6.1 Selectivity
Immunoassays possess moderate discriminatory ability.
Commonly
encountered drugs should be tested prior to using the assays to test the
reactivity with the assay.
13.6.2 Matrix
Effects
Solvents, pH, light, and temperature can interfere with
the reaction.
A
study should document controls placed in procedure to reduce or eliminate
adverse effects
13.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
13.6.4 Accuracy
13.6.4.1 Precision (Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum of 10 times.
13.6.4.2 Trueness
Trueness must be determined for quantitative methods.
13.6.5 Range
13.6.5.1 Limit of detection (
Immunoassays have sufficient sensitivity to detect drugs
in the nanogram level.
Using
various concentrations of known standards and measuring the response should
determine the limit of detection.
13.6.5.2 Limit of quantitation (LOQ)
13.6.5.3 Linearity
13.6.6 Robustness
Determine
the amount of change to instrumental parameters that will still allow for a
suitable comparison.
13.6.7 Ruggedness
13.7 Uncertainty
13.8 Quality
Control
13.9 Reference
14.1 Technique
Strengths
Selectivity: Mixed-melting determination adds selectivity
by first running the sample alone, then mixing a standard of the suspected
compound with the sample and checking for agreement.
14.2 Technique
Limitations
The
temperature should rise at a constant, slow rate to allow for accurate
observation.
Samples should be dry and free from diluents or other
adulterants.
Re-crystallization of street samples may be necessary.
Availability may be limited.
14.3 Purpose/Scope
Melting
point determination is the determination of a physical property of a compound
that may be compared to literature values or a standard.
This
also can be used to aid in the identification of a compound when a mixed
melting point determination is performed.
14.4 Analytical
Method
14.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment
14.4.2 Instrumental
parameters
Identify the instrument and equipment utilized.
Identify the procedures to be utilized.
List
instrumental conditions such as temperature rate increase and melting range.
14.5 Reference
Materials
Reference
materials of the drugs to be analyzed are suitable for method validation.
14.6 Performance
Characteristics
14.6.1 Selectivity
Melting
point ranges provide physical information about the analyte.
Selectivity
is greatly increased by utilizing the mixed-melting point technique.
Standards
should be run with each set of samples.
For
determination of closely related compounds, standards of each should be tested
on the system to show selectivity.
14.6.2 Matrix
effects
14.6.3 Recovery
14.6.4 Accuracy
14.6.4.1 Precision (Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
14.6.4.2 Trueness
14.6.5 Range
14.6.5.1 Limit of detection (
Melting
point determination requires sufficient sample for the apparatus being
employed.
Determine
the sensitivity by measuring the response of different amounts of analyte.
For most systems this is in the milligram range.
14.6.5.2 Limit of quantitation (LOQ)
14.6.5.3 Linearity
14.6.6 Robustness
14.6.7 Ruggedness
14.7 Uncertainty
14.8 Quality
Control
14.9 Reference
14.9.1 The
15 ULTRAVIOLET
SPECTROPHOTOMETRY (UV)
15.1 Technique
Strengths
Samples can be recovered for additional tests.
UV can easily be combined with HPLC for greater
selectivity and specificity
Hyphenation with chromatography also makes automation of
the technique easy.
15.2 Technique
Limitations
Compounds lacking suitable chromophore provide no signal.
High
analyte concentration in the sample may cause full absorption at all
wavelengths yielding saturated spectra.
UV spectrum often varies depending upon the pH of the
sample solution.
Chemical composition may change during the analysis.
15.3 Purpose/Scope
UV
yields rough structural information providing modest selectivity to allow for
some discriminating capability
It
can be used as a quantitative method.
Moreover,
it is more commonly used in combination with liquid chromatography for greater
selectivity.
15.4 Analytical
Method
15.4.1 Sample
preparation
List
the required sample preparation schemes and the introduction techniques for the
selected instrument and equipment
15.4.2 Instrumental
parameters
Identify
the instrument and equipment utilized, to include the UV spectrophotometer (or
detector) used in the present laboratory.
List instrumental conditions.
15.4.3 Calculations
The
equations and calculations used in quantitation must be delineated to include
unit specifications, number of repeated measurements, significant figures,
conditions for data rejection, reference values and uncertainty determination.
15.5 Reference
Materials
Utilize
a Holmium Oxide filter for conducting a validation run on the UV and compare to
the reference spectrum provided.
Repeat
this process utilizing commonly encountered drugs suitable for this method.
15.6 Performance
Characteristics
15.6.1 Selectivity
A
Ultra-Violet Spectrum yields limited structural information and providing
modest selectivity to allow for some discriminating capability.
Validation
data will show the ability of the method to discriminate between different
compounds.
Standard
spectra collection (library) shall be used as reference in the identification
of the active compound.
15.6.2 Matrix
effects
Organic
solvents have varying UV absorbance and this may interfere with the absorbance
of the analytes.
Influence
of sample preparation on the results (direct dissolving of the sample in
buffer/solvents, liquid-liquid extraction) should be investigated with extreme
care, taking in the consideration the pH, type of buffer and solvent, and the
matrix.
15.6.3 Recovery
Sample recovery may be determined for quantitative
analysis.
15.6.4 Accuracy
15.6.4.1 Precision (Repeatability/Reproducibility)
Demonstrate
the reproducibility of the instrument by running a reference material a minimum
of 10 times.
15.6.4.2 Trueness
Trueness must be determined for quantitative methods.
15.6.5 Range
15.6.5.1 Limit of detection (
Determine the
15.6.5.2 Limit of quantitation (LOQ)
Determine
the lowest concentration that has an acceptable level of uncertainty.
The
LOQ is the lower end of the linear determination.
15.6.5.3 Linearity
Determine
the mathematical relationship (calibration curve) that exists between
concentration and response over a selected range of concentration. For purposes of UV this is normally a
straight line.
The
highest linear concentration serves as the upper limit for quantitative purposes.
15.6.6 Robustness
Determine
the amount of change in instrumental parameters that will still allow for the
level of acceptance required (e.g. vary solvent, pH, scan time, analysts,
etc.).
15.6.7 Ruggedness
15.7 Uncertainty
Uncertainty should be evaluated for quantitative methods.
15.8 Quality
Control
15.9 Reference
15.9.1 The
Fitness for Purpose of Analytical Methods, Eurachem Guide, Dec 1998, p43 and
49.
Section II: A Completed Validation
Plan
The following demonstrates a purpose-defined
validation plan for a particular method within an individual laboratory. The aim is to show how a complete validation
plan may appear.
The
following example should not be directly applied to methodology used by any
laboratory without first considering the specific purpose of a method and its
relevant operational environment.
Scenario: Laboratory x is tasked with
validating a qualitative and quantitative method for the analysis of
heroin. The laboratory has defined the
performance specifications necessary to achieve the laboratory standards and
meet the customer requirements. The
method to be validated utilizes GC/MS.
Validation
plan for GC/MS identification and quantitation of heroin
1 Purpose/Scope
Establish if the GC/MS method for heroin
identification and quantitation meets the laboratory and customer
specifications (performance specifications) by examination and review of
objective evidence.
Performance
specifications
-
Selectivity: Sufficient to enable full separation of heroin from other
opiates
-
Matrix effects: Sufficient to enable full separation of heroin from
diluents and cutting agents typically met in street samples
-
Recovery: > 95%
-
Precision (Repeatability): < 5%
-
Precision (Reproducibility): < 8%
-
Trueness: < 8%
-
Limit of detection (
-
Limit of quantitation (LOQ): 1.0 %
-
Linearity: correlation coefficient > 0.99
-
Robustness: sufficient for routine work
-
Ruggedness: not significant as the method will be applied only in one
laboratory
-
Uncertainty: expanded uncertainty < 10%
Process
review
The results of the validation experiments will be
reviewed against the performance specifications:
A. If the
experimental results achieve the performance specifications, the method is
validated (fit for purpose).
B. If the
experimental results do not achieve the performance specifications, the
following options will be considered:
1. The performance specifications may reviewed,
amended if appropriate and validation accepted based on redefined
specifications.
2. The method will be redeveloped and
revalidated.
2 Analytical Method
Sample
preparation
Approximately
20mg (15 60 mg) of homogenized sample material is accurately weighed in a
test tube and the weight recorded. The
powder is dissolved in 5.0 mL of methylene chloride containing the internal
standard (5a-cholestane, 0.5 mg/mL). The test tubes are capped,
shaken for 15min and centrifuged at 2500 rpm for 5min. The mixture is filtered if turbid. Approximately 1mL of the methylene chloride
solution is transferred into a GC vial for the GC/MS analysis. The vial is capped and its tightness
checked.
Stability
of analyte
Stability
of heroin in methylene chloride will be evaluated during the course of the
validation. Additionally, shelf life of
the heroin standard solutions will be investigated by storing the solutions for
three weeks and by controlling their concentrations against the calibration
curve prepared on day one.
Instrumental
parameters
Instrument: HP
6890 gas chromatograph with autosampler, HP 5973 mass selective detector,
Agilent MS Chemstation rev. B.01.00 (Agilent technologies).
Column: A
5% phenyl methyl silicone capillary columns (HP-5MS), 30 m (L) x 0.25 mm
(i.d.), df 0.25 mm (Agilent technologies).
Carrier gas: Helium,
25 cm/s at 150 °C, constant flow
Sample introduction: 1 mL split, 60 mL/min total flow, 1:148 split ratio (gas saver 20mL/min
after 1.5 min), glass wool packed liner with a volume of 990 ml.
Temperatures:
Injector: 250 °C
Oven T-program: 150
°C, 10 °C/min, 300 °C (10 min)
GC/MS interface: 310 °C
MS information:
Solvent delay: 2.5
min
Mass range: 30
- 550 a.m.u.
Sample rate #: 2,
A/D samples 4
MS quad temp: 150
°C
MS source: 230
°C
The total ion chromatogram is used for quantitation.
Calculations
A single-point calibration based on the internal
standard method is utilized.
3 Reference Materials
MS reference material - PFTBA is used to calibrate
the MS in accordance with the procedures recommended by the instrument
manufacturer.
Drug reference material - Certified heroin
hydrochloride monohydrate, 98.83 %, M-29-HC-500 purchased from
Internal standard reference material - 5a-cholestane,
p.a., purchased from Merck,
4 Performance
Characteristics
4.1 Selectivity
Selectivity
of the GC/MS screening method will be investigated by adding known substances,
i.e. other opiates and commonly encountered street drugs. Difference in retention time will be used as
the first criterion and the mass spectrum as the second criterion to estimate
selectivity.
4.2 Matrix
effects
The influence of cutting agents and adulterants
commonly encountered in street drugs will be investigated. Paracetamol
(acetaminophen), a known interfering substance, will be added to 10 different
known heroin samples (n=10) and the analyses carried out. Other commonly encountered cutting agents and
adulterants will be evaluated in the same manner.
Matrix effects may also have great impact on
repeatability. This will be investigated
by analyzing six different heroin street samples. Ten replicates of each sample will be prepared
and analyzed and their concentrations calculated. Repeatability in terms of RSD will be derived
from the concentrations.
This
data will be used to evaluate the magnitude of the effect, if any, on trueness
and precision.
4.3 Recovery
As
the method is based on dissolution, recovery can be evaluated using only
typical cutting agents, i.e. lactose.
Recovery of the method will be investigated by standard addition
technique. Heroin standard will be mixed
with lactose in different ratios (n=10) covering a concentration range of 5
100%. The sample will be prepared and
analyzed as described above. Heroin
concentration will be measured against the calibration curve and recovery
established.
4.4 Accuracy
4.4.1 Precision (repeatability)
Qualitative analysis
Ten
preparations of one heroin street sample will be analyzed in random order. Correct identification of heroin will be
evaluated against laboratory requirements.
Quantitative analysis
Within
day repeatability will be investigated by analyzing ten replicates of one
heroin street sample. Repeatability in
terms of RSD will be derived from the quantitative results.
4.4.2 Precision (reproducibility)
Qualitative analysis
Reproducibility
will be investigated by analyzing a sample ten times by two different operators
and on two different instruments and the results compared.
Quantitative analysis
Reproducibility
will be investigated by two operators independently analyzing one sample ten
times. Student t-test of means will be
applied to the data with a 95% confidence interval.
4.4.3 Trueness
Trueness
will be investigated by comparing values obtained by the current method with
those obtained by a validated reference method (HPLC). Trueness will be investigated by analyzing a
heroin street sample ten times. Trueness
will be estimated through the Student t-test of means. Statistical difference between the results of
the GC/MS and the reference method will be considered systematic error.
4.5 Range
4.5.1 Limit of detection (
4.5.2 Limit of quantitation (LOQ)
LOQ
will be investigated through the analysis of ten replicates. The concentration of the replicates is
determined by a serial dilution of a known heroin solution (low end
concentrations). LOQ will be defined as
the lowest concentration in which relative standard deviation (RSD) is not
higher than the repeatability of the method.
4.5.3 Linearity
Linearity
will be investigated using a minimum of eight different concentrations (0.1,
0.5, 1.0, 5.0, 10, 50, 80 and 100 % of heroin).
The peak areas will be plotted against concentrations, a calibration
curve drawn and correlation coefficient calculated.
4.6 Robustness
Robustness will be investigated by varying
analytical procedure, i.e. the volume of the methylene chloride and the weighed
amount of heroin. Evaluation of other
parameters is not considered relevant for this application.
4.7 Ruggedness
Ruggedness is not relevant in this study as the
method will be utilized in one laboratory only and the reproducibility study
includes evaluation of intra-laboratory variation.
4.8 Uncertainty
Uncertainty will be calculated taking into
consideration both systematic error (trueness) and the random error
(repeatability/reproducibility). The
expanded uncertainty will be established using a 95% confidence level.
5 Quality Control
Acceptance
criteria for quality control parameters will be established on the basis of the
validation results prior to implementation of the method.
6 References
The testing laboratory will provide the selected
supporting reference literature relating to the validated method.
End of Document